Stopwatch timer operation, If an airflow alert activates, Resetting the airflow alert system – Labconco Total Exhaust Biological Safety Cabinets 34368 Series User Manual

Page 46: Working in the biosafety cabinet

Advertising
background image

Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

40

Original instructions

4. When the timer reaches 0:00:00, an audible alarm will sound, and the

timer will reset itself and repeat the countdown.

5. Press the

OK

button to pause the timer.

6. Press the

Menu

button to clear the interval timer and return to the

main timer menu.

Stopwatch Timer Operation

1. When selected, the Stopwatch Timer menu is displayed on the LCD.

The timer defaults to 0:00:00.

2. Press the

OK

button to start the timer.

3. Press the

OK

button again to zero the timer.

4. Press the

Menu

button to clear the stopwatch timer and return to the

main timer menu.

If An Airflow Alert Activates

The most common causes of an Airflow Alert are:

• Blockage of the inlet grilles or exhaust outlet.

• Removal of the work surface or grille during operation.

Resetting the Airflow Alert System

The Airflow Alert automatically resets to normal operation once the motor
speed has stabilized.

Working In the Biosafety Cabinet

Note: A more thorough review of using the BSC can be found in: Biosafety
in Microbiological and Biomedical Laboratories (BMBL)
, Published
by the Centers for Disease Control and Prevention
(

www.cdc.gov/biosafety/publications

).


Planning

• Thoroughly understand procedures and equipment required before

beginning work.

• Arrange for minimal disruptions, such as room traffic or entry into the

room while the cabinet is in use.