Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

42

Original instructions

Work Techniques

• Keep all materials at least 4 inches (100 mm) inside from the sash,

and perform all contaminated operations as far to the rear of the work
area as possible.

• Segregate all clean and contaminated materials in the work area.

• Arrange materials to minimize the movement of contaminated

materials into clean areas.

• Keep all discarded contaminated material to the rear of the work area.

• Avoid moving materials or the operator's hands and arms through the

front access opening during use.

• Avoid the use of an open flame. Use disposable labware or an electric

incinerator as alternatives.

• Use proper aseptic technique.
• Avoid using techniques or procedures that disrupt the airflow patterns

of the cabinet.

• If there is a spill or splatter during use, all objects in the cabinet

should be surface decontaminated before removal. Thoroughly
disinfect the working area of the cabinet WHILE IT IS STILL IN
OPERATION, to prevent the release of contaminants from the
cabinet.

Final Purging

• Upon completion of work, the cabinet should be allowed to operate

for two to three minutes undisturbed, to purge airborne contaminants
from the work area.

Unloading Materials and Equipment

• Objects in contact with contaminated material should be surface

decontaminated before removal from the cabinet.

• All open trays or containers should be covered before being removed

from the cabinet.

Wipe-Down

• Wipe down the interior surfaces of the cabinet with a suitable

disinfectant, or 70% ethanol, and allow to dry.

• Periodically lift the work surface and wipe down the area beneath it.