Labconco Total Exhaust Biological Safety Cabinets 34368 Series User Manual

Page 58

Advertising
background image

Chapter 7: Troubleshooting

Product Service 1-800-522-7658

52

Original instructions

PROBLEM

CAUSE

CORRECTIVE ACTION

Airflow Alert goes
off and/or there is a
slight decrease in
filter life remaining
gauge

HEPA filter loading

The gauge reading steadily decreases as
the cabinet is used.

Blockage of the
return air slots or
grille

Check all return air slots and grilles to
ensure that they are not blocked or
restricted.

Blockage of the
exhaust outlet

Ensure that the exhaust outlet is not
blocked or restricted.

Blockage

or

restriction under the
work surface

Ensure that the towel catch and plenum
beneath the work surface are
unobstructed.

Exhaust Error goes
off and alarm
sounds 15 seconds
after the blower is
started

Roof blower is off

Ensure that roof blower is working.

Exhaust damper is
improperly set.

Ensure that the exhaust damper is
properly set by a qualified certifier.

Remote blower is
improperly sized

Confirm that the roof mounted blower
meets the volume and pressure
requirements of the cabinet.

Mechanical failure of
the exhaust system

Inspect the exhaust system.

Obstruction in the
exhaust system

Inspect the exhaust system.

Leak or additional
‘load’ on the exhaust
system

Inspect the exhaust system.

Loading

of

the

exhaust HEPA filter

Replace the Exhaust HEPA filter.