Labconco Total Exhaust Biological Safety Cabinets 34368 Series User Manual

Page 47

Advertising
background image

Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

41

Original instructions

Start-up

• Turn off UV light if included.

• Slowly raise the sash until the bottom of the sash aligns with the sash

indicator decal located on the left side of the work area.

• Turn on the fluorescent light and cabinet blower if the Smart-Start

features have not been activated.

• Check the air grilles for obstructions.

• Allow the cabinet to operate until the display screen is shown.

Warming

up

Screen Display

Screen

• Wash hands and arms thoroughly with germicidal soap.

• Wear appropriate personnel protective equipment (PPE).

Wipe-Down

• Raise the sash to its full open position (approximately 21.75 inches or

552 mm). Mute the alarm by depressing the "OK/Mute" switch.

• Wipe down the interior surfaces of the cabinet with 70% ethanol, or a

suitable disinfectant, and allow to dry.

Loading Materials and Equipment

• Only load the materials required for the procedure. Do not overload

the cabinet.

• Do not obstruct the front, side, or rear return air grilles.

• Large objects should not be placed close together.

• Slowly close the sash until it is in the correct operating position.

• After loading the cabinet, wait two to three minutes to purge airborne

contaminants from the work area.