Work Surface and HEPA Filtered Enclosure Installation

1. Level the base cabinets and the work surface. Work surface should be

placed flush with the front of the base cabinet as shown in Figure 3-1.

2. Position the work surface in its intended location and with the front of the

work surface towards you. (Rear mounting holes are located close to the
rear edge.)

3. Secure the work surface to the base cabinet with a structural adhesive or

silicone sealant.

4. Insert the supplied mounting screws in the four holes. Allow a minimum

of 1/8" clearance under the head of the screw for positioning the enclosure.

5. Place the enclosure on the work surface and slide the rear flange and front

air foil flanges under the mounting screw heads.

6. Tighten the four screws to complete the installation.