Chapter 8: Troubleshooting and Servicer Operating Log

Product Service 1-800-522-7658

57

PROBLEM

CAUSE

CORRECTIVE ACTION

Blower won’t
operate.

Unit not plugged
into outlet.

Plug the enclosure into appropriate
electrical service.

Circuit breaker(s)
or Ground Fault
Interrupter.

Reset circuit breaker.

Blower wiring is
disconnected.

Inspect blower wiring.

Blower switch is
defective.

Replace switch.

Motorized impeller
or blower is
defective.

Replace motorized impeller or blower.
See Chapter 6.

Low face velocity
or poor
containment of
contaminants.

Enclosure sash not
closed.

Close sash to the lowest position.

HEPA filter
clogged.

Replace HEPA filter or increase speed.

Carbon filter loaded
with chemicals.

Replace carbon filter.

Blower and lights
won’t operate.

Unit not plugged
into outlet.

Plug enclosure into appropriate
electrical service.

Circuit breaker(s)
tripped.

Reset or replace circuit breaker.

Light does not
work.

Lamp not installed
properly.

Inspect lamp installation.

Lamp wiring
disconnected.

Inspect lamp wiring.

Defective lamp.

Replace lamp.

Light switch is
defective.

Replace light switch.

Defective electronic
ballast.

Replace ballast.

UV closure panel
not installed .

Install UV closure panel (UV lights
only).

Magnetic switch for
UV closure panel
defective.

Replace switch (UV lights only).

UV timer circuit
board defective.

Replace circuit board (UV lights only).