Speed control replacement, Trace odor carbon filter installation – Labconco Class I Enclosure Models 3980423 User Manual

Page 53

Advertising
background image

Chapter 6: Maintaining Your Filtered Enclosure

Product Service 1-800-522-7658

49

Speed Control Replacement

1. Remove the bracket that the speed control is attached to. See Figures 6-1, and

6-2.

2. Remove the two screws holding the speed control using a Phillips screwdriver.

Refer to Appendix A for Replacement Parts Diagram.

3. Disconnect all wires leading to the speed control. Connect wires on new speed

control in the same position as the old speed control.

4. Reassemble to the system in the same position and with the same screws that

were removed earlier.

Trace Odor Carbon Filter Installation

In the event that low volumes of solvents are required, a Trace Odor Carbon Filter
can be installed.

1. Remove the screws (2) located on each side of the upper diffuser.
2. Lift and remove the upper diffuser.
3. Place the Trace Odor Carbon Filter on top of the HEPA filter with the gasket-

side DOWN. Use care when maneuvering past the airflow sensor. Be careful
not to harm the HEPA filter media that is exposed.

4. Replace the upper diffuser
5. Replace all screws.

Upper Diffuser

Trace Odor
Carbon Filter

Airflow Sensor

Advertising