Hepa filter leak test, Hepa – Labconco Class I Enclosure Models 3980423 User Manual

Page 42

Advertising
background image

Chapter 6: Maintaining Your Filtered Enclosure

Product Service 1-800-522-7658

38

38

HEPA Filter Leak Test

Purpose

After installing the new HEPA filter, the HEPA filter should be leak checked.
This test is performed to determine the integrity of the HEPA filter, the filter
housing, and the filter mounting frames. Leak testing is to be done by a qualified
technician with calibrated equipment. Remove the top perforated exhaust cover by
using a Phillips screwdriver to unfasten the (2) screws used to secure (2) clips. See
Figure 6-1. The filter passes the leak test at .01% or better. Reference Leak
Testing and Photometer scanning from the Institute of Environmental Services
(IES-RP-CC001.3)

Apparatus

1. An aerosol photometer ATI model 2D, 2E, 2G or equivalent. Air

Techniques Hamilton Associates inc. 11403 Cron Ridge Dr. Owings Mills,
MD 21117

2. One aerosol generator of the Laskin nozzle(s) type. An aerosol of mineral

oil or suitable liquid shall be created by flowing air through it. The compressed
air supplied to the generator should be adjusted to a pressure of 10± 1 psig.
during operation. Air Techniques Inc. Model TDA-4A or equal. One nozzle at
10 psig is (67.5 cfm x 100ug/l)/(Vol. of air), For the 2' at 90 fpm or 130 cfm,
one nozzle @10 psig is 13,500/130 cfm = 52 ug/l. For the 3' at 90 fpm or 200
cfm, one nozzle @ 10 psig is 34 ug/l. For the 4' at 90 fpm or 265 cfm, one
nozzle at 10 psig is 26 ug/l.

3. Mineral oil (Catalog #1491400).
4. Sampling Nozzle, Rectangular 1/2" x 3-1/4", Air Techniques, Inc.

Procedure

For the ATI 2G Photometer

1. Turn on the photometer and allow it to operate for a minimum of 5 minutes.

Leave the valve in the “CLEAR” setting.

2. Press the “ENTER” keypad. Press the “REF” keypad.

The display will display “P1” for approximately 1 second, and then display
a numerical value.

3. The display will display “P1” for approximately 1 second, and then display a

numerical value.

4. Using the “^” or “ν” keypads, respectively, increase or decrease the numerical

value until it equals 52 (2'), 34 (3'). 26 (4') for the enclosure at 90 fpm inflow
velocity.

5. Press the “ENTER” Keypad. The photometer will scan for 15 seconds, and

then the “0” keypad will flash. Press the “Enter” keypad. The unit will scan for
5 seconds, the display will read “0000,” and the unit will sound a confirming
tone.

HEPA

Advertising