Labconco Class I Enclosure Models 3980423 User Manual

Page 43

Advertising
background image

Chapter 6: Maintaining Your Filtered Enclosure

Product Service 1-800-522-7658

39

6. Set the valve to “DOWNSTREAM.” Place the palm of your hand over the

sampling port of the pistol. There should be a strong vacuum at this port. If the
vacuum is weak, contact Air Techniques Hamilton Associates.

7. Turn the enclosure on and let it operate for a minimum of 5 minutes.

8. If necessary, adjust the speed control of the enclosure to maintain the following

airflows at 90 fpm; 2' (130cfm), 3' (200 cfm, 4' (265 cfm.)

9. Position the aerosol generator discharge in the intake of the baffle inside the

enclosure.

10. Start the aerosol generator (Pressure to be +/- 1 PSIG). Ensure that one Laskin

nozzle is in the “open” position.

11. Allow the generator to operate for a minimum of 15 seconds. For all integral

motorized impeller models, scan the downstream exhaust side of the HEPA
filter by passing the sampling nozzle of the gun in slightly overlapping strokes
over the entire surface the filter, with the sampling port not more than 1 inch
from the surface of the filter media. Scan the entire periphery of the filter and
the gasket between the filter frame and the enclosure frame. Scanning shall be
done at a traverse rate of not more than 2 inches per second.

NOTE: For the Purifier Enclosures ducted to the outside, place the

sampling nozzle in the center of the remote blower exhaust.

Acceptance

Aerosol penetration shall not exceed 0.01 percent measured by the photometer.

Setting the Inflow Face Velocity with the

Speed Control Adjustment

1. Remove the front panel by loosening the (2) Phillips screws on top that

secure the front panel.

2. The speed control is located on the electrical subassembly located behind

the switched control panel and below the front panel. See Figure 6-1.

3. Adjust the speed control with a small Phillips screwdriver by turning the

screw counterclockwise to increase blower speed or clockwise to decrease
the blower speed. The speed control is very sensitive, so proceed with
caution.

4. Measure the inflow velocity per the averaging technique outlined in

Chapter 3 and adjust the speed control slowly for the desired speed. Allow
the speed to stabilize and re-measure the inflow velocity to confirm.

5. Replace the front panel and tighten the screws.

Advertising