Location and air current requirements, Airflow specifications – Labconco Purifie Horizontal Clean Benches 3970424 User Manual
Page 10

Chapter 2: Prerequisites
Product Service 1-800-522-7658
6
6
Location and Air Current Requirements
Purifier Vertical Clean Benches have been designed to provide particulate-free
clean air meeting Class 5 conditions by negating typical cross drafts and
turbulence within the opening. However, as a precautionary safety measure and a
higher level of quality management, it is recommended that the clean bench be
placed in an area away from:
High traffic areas where walking might cause an air disturbance or be a
nuisance.
Overhead or wall HVAC diffusers, fans, radiators or other lab equipment
producing air currents.
Doorways or windows that may be opened.
Do not position the unit so that it is difficult to operate the main
disconnect device.
Ne placez pas l'appareil de sorte qu'il est difficile de faire
fonctionner le dispositif principal de déconnexion.
Airflow Specifications
Purifier Vertical Clean Benches use an integral motorized impeller to draw room
air from the top and through the enclosure. This room air is then pushed down
through the HEPA filter. The enclosure is factory set to deliver 40-55 fpm
downflow velocity measured 6" below the upper diffuser. The Class 5 HEPA-
filtered air flows across the work area and exits out the front of the enclosure.
Data for the outflow exhaust volume, downflow velocities and out-flow face
velocities are listed for each Vertical Clean Bench model below.
Enclosure
Width
Model Description
Downflow
Velocity
(fpm)
Out-
Flow
Exhaust
Volume
(CFM)
Noise
Pressure
db(A)
Average
Outflow
Face
Velocity
(fpm)
2'
2' Vertical Clean Bench
30
40
50
60
120
165
210
250
53-57
58-60
60-61
62-63
85
115
145
170
3'
3' Vertical Clean Bench
30
40
50
60
180
250
310
370
58-60
60-62
62-66
66-70
85
115
145
170
4'
4' Vertical Clean Bench
30
40
50
60
240
330
420
500
58-62
60-64
62-66
66-70
85
115
145
170
!
- Purifie Horizontal Clean Benches 3970423 Purifie Horizontal Clean Benches 3970421 Purifie Horizontal Clean Benches 3970420 Purifie Horizontal Clean Benches 3970404 Purifie Horizontal Clean Benches 3970403 Purifie Horizontal Clean Benches 3970401 Purifie Horizontal Clean Benches 3970400 Purifie Horizontal Clean Benches 3970324 Purifie Horizontal Clean Benches 3970323 Purifie Horizontal Clean Benches 3970321 Purifie Horizontal Clean Benches 3970320 Purifie Horizontal Clean Benches 3970304 Purifie Horizontal Clean Benches 3970303 Purifie Horizontal Clean Benches 3970301 Purifie Horizontal Clean Benches 3970300 Purifie Horizontal Clean Benches 3970224 Purifie Horizontal Clean Benches 3970223 Purifie Horizontal Clean Benches 3970221 Purifie Horizontal Clean Benches 3970220 Purifie Horizontal Clean Benches 3970204 Purifie Horizontal Clean Benches 3970203 Purifie Horizontal Clean Benches 3970201 Purifie Horizontal Clean Benches 3970200