Hapter, Roubleshooting – Labconco FreeZone 2.5 Liter Freeze Dry Systems 76705 Series User Manual

Page 36

Advertising
background image

Product Service 1-800-522-7658

32










C

HAPTER

7

T

ROUBLESHOOTING


Refer to the following if your Freeze Dryer fails to operate properly. If the
suggested corrective actions do not solve your problem, contact Labconco for
additional assistance.

Labconco Freeze Dry Systems that are clean, dry and without samples attached
should reach a vacuum of 0.133 mBar within 10 minutes and should achieve an
ultimate vacuum of 0.040 mBar within 18 hours when the refrigeration is
operating. If the Freeze Dryer does not obtain a satisfactory vacuum, perform the
following maintenance tests.

I. Vacuum Pump


First make sure that the vacuum pump operates. If it fails to operate, check the
electrical connections of the Freeze Dryer to the power source and then check the
electrical connection of the vacuum pump to the Freeze Dryer. If the vacuum
pump has a power switch, make sure that it is turned on. If the vacuum is not
adequate when the vacuum pump is operating, proceed with the following steps:

1.

Check the oil and ensure it is clear and clean. If the oil looks cloudy or has

any particulates, replace the pump oil. Sometimes it may be necessary to
flush the pump with clean oil several times. To flush the pump, run the
pump 5 to 10 minutes to allow the oil to warm up. Drain the oil and refill
with clean oil. Repeat as necessary.

2.

Check the oil level in the pump. Ensure it is filled to the correct level.

3.

Check vacuum hose connections from the pump to the Freeze Dryer and try

running the unit.


If vacuum problems continue, consider obtaining a second vacuum gauge capable
of reading a vacuum of 0.010 mBar. It is often useful in determining if the
vacuum pump is operating properly and the vacuum sensor reading is accurate.