Appendix C: Freeze Dryer Specifications

Product Service 816-333-8811 or 1-800-522-7658

52

Catalog #

Voltage

Frequency

Phase

Amperage

Nominal

Operating

Range

Nominal

w/o

Vacuum

Pump

Max. with

Vacuum

Pump

7420020

115V

103-127

60

1

10.0

16

100V

90-110

50

1

10.0

16

7420021

115V

103-127

60

1

10.0

16

100V

90-110

50

1

10.0

16

7420030

230V

198-254

50

1

5.1

10

7420031

230V

198-254

50

1

5.1

10

7420040

230V

187-253

60

1

5.1

10

7420041

230V

187-253

60

1

5.1

10

7420060

230V

198-254

50

1

5.1

10

7420061

230V

198-254

50

1

5.1

10

7420070

230V

198-254

50

1

5.1

10

7420071

230V

198-254

50

1

5.1

10

Environmental Conditions

Indoor use only.

Maximum altitude: 6562 feet (2000 meters).

Ambient temperature range: 41° to 104°F (5° to 40°C).

Maximum relative humidity: 80% for temperatures up to 88°F (31°C),

decreasing linearly to 50% relative humidity at 104°F (40°C).

Main supply voltage fluctuations not to exceed ±10% of the nominal

voltage.

Transient overvoltages according to Installation Categories II (Overvoltage

Categories per IEC 1010). Temporary voltage spikes on the AC input line
that may be as high as 1500V for 115V models and 2500V for 230V
models are allowed.

Used in an environment of Pollution degrees 2 (i.e., where normally only

non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.