INSTALLATION

13

4. Hold the bubbler straight (with the vertical tubes perpendicular to the glove box base) and

tighten the top hanger bracket #8 x 1/2" screw. Secure the bubbler to the hanger bracket by
tightening the _-20 nut.

5. Remove the pipe plug from the fill spout and fill the bubbler with vacuum pump oil (.87

specific gravity) to the desired pressure relief level (must be less than 10) as indicated on
the graduated sight glass (inches H

2

O). The recommended level is 7 inches for most

applications. When nearing the desired fill level, add a small amount of oil at a time and
wait for it to drain into the bubbler and sight glass. When the desired level is reached,
install and tighten the fill spout plug.

NOTE: When loosening or tightening the fill plug or drain cock, be sure to hold the
bubbler’s stationary fitting with a wrench to avoid damage to the bubbler’s soldered joints.

Figure 5