ROUTINE MAINTENANCE

32

• Stretch the new glove’s beaded cuff over the old glove cuff and into the groove nearest the

window.

• To remove the old glove, grasp the new glove surface and manipulate the old glove bead

free from the outer groove.

• With the other glove hand inside the box, pull the old glove into the glove box.

• Install the O-ring over the new glove onto the outer groove.

Fluorescent Lamp Replacement

• Turn the light switch off and unplug the lamp cord from the receptacle located on the

outside top of the cabinet.

• From the top of the lamp housing, remove the two Phillips-type screws attaching the lamp

assembly to the top edge of the window frame.

• Remove the lamp assembly from the unit.

• The lamp tube can then be easily removed and replaced. Replacement lamp part number

12779, fluorescent lamp #F30T12/CW/RS may be purchased from most local lighting
supply stores.

• Re-install the lamp assembly and attach with the two Phillips-type screws. Plug in the

power cord and switch on the lamp. Refer to the illustration below.

Figure 17