ROUTINE MAINTENANCE
32
• Stretch the new glove’s beaded cuff over the old glove cuff and into the groove nearest the
window.
• To remove the old glove, grasp the new glove surface and manipulate the old glove bead
free from the outer groove.
• With the other glove hand inside the box, pull the old glove into the glove box.
• Install the O-ring over the new glove onto the outer groove.
Fluorescent Lamp Replacement
• Turn the light switch off and unplug the lamp cord from the receptacle located on the
outside top of the cabinet.
• From the top of the lamp housing, remove the two Phillips-type screws attaching the lamp
assembly to the top edge of the window frame.
• Remove the lamp assembly from the unit.
• The lamp tube can then be easily removed and replaced. Replacement lamp part number
12779, fluorescent lamp #F30T12/CW/RS may be purchased from most local lighting
supply stores.
• Re-install the lamp assembly and attach with the two Phillips-type screws. Plug in the
power cord and switch on the lamp. Refer to the illustration below.
Figure 17