Chapter 5: Using Your Glove Box

Product Service 1-800-522-7658

22

Loading Materials and Equipment (see Figure 5-1)

• Load only the materials required for the procedure. Do not overload

the glove box.

• Be aware that gloves will inflate when the glove box is operated

under negative pressure. Ensure items in the glove box will not
obstruct the movement of the gloves.

• After loading, wait 2-5 minutes at 50-100% blower speed to purge

airborne contaminants from the work area before beginning
procedures.

Work Techniques

• Segregate all clean and contaminated materials in the work area.

Final Purging (see Figure 5-1)

• All open trays, weigh vessels or containers must be sealed before

being removed from the glove box.

• Objects in contact with contaminated material must be surface

decontaminated before removal from the glove box.

• The interior surfaces should be cleaned per user’s protocol.
• Upon completion of work, the glove box should be allowed to operate

for 5 minutes undisturbed at 100% speed or 10 minutes at 50% speed
to purge and dilute airborne contaminants from the work area.

• After purging the main chamber, the transfer chamber should also be

purged for 5 minutes at 100% speed or 10 minutes at 50% speed with
the inner door open, before removing sealed material.

Shutdown

• Turn off the exhaust system and light.

• Keep outer door closed when not in use.