Chapter 2: Prerequisites

Product Service 1-800-522-7658

3

Support, Vibration and Preventive Requirements

In the preparation of a glove box site, please consider the following:

• A bench or stand that is rigidly mounted to the floor or fixed to the wall,

but not both, may be appropriate. 35" to 40" (889mm-1016mm) is
typical for standing height. (Labconco stands offered in Chapter 7 vary
from 33" to 40".)

• The corners of a building typically have less vibration than the center,

which promotes analytical balance stability.

• The bench typically should not contain any vibration-producing

equipment, such as shakers or pumps.

• A marble slab with dampening pads placed within the enclosure is an

effective low cost means of controlling vibration (see Chapter 7:
Accessorizing Your Filtered Glove Box
).

Location Requirements

The Precise HEPA-Filtered Glove Boxes have been designed to rest on a
typical 29"-30" (737mm-762mm) deep work surface. The height should be
35"-40" for standing position. Avoid placing the glove box in high traffic
areas where walking might disrupt the operator or experimentation.

Exhaust and Blower Requirements

The Precise HEPA-Filtered Glove Box, as manufactured, uses an integral
motorized impeller to pull room air through the inlet and exhaust HEPA
filters and discharge the clean air back into the laboratory. This negative
pressure operation mode pulls all internally-contaminated air through the
exhaust HEPA filter. The HEPA-filtered exhaust air is then forced out the
top of the glove box. An optional FilterMate™ Portable Exhauster with a
carbon filter may be installed downstream of the exhaust to adsorb low levels
of organic fumes, formaldehyde, or ammonia gases.

The HEPA-filtered exhaust air from the glove box can be exhausted to the
outside with the installation of a thimble exhaust connection kit and remote
blower listed in Chapter 7. As an option, if necessary, the built-in motorized
impeller can be wall mounted if desired.

Labconco does not recommend a hard duct connection using the built-in
motorized impeller as the source of airflow out of the facility. This
creates a positive pressure in the exhaust duct.