7. Move the cursor to the ‘TMR’ position
8. Select the amount of time (0 thru 9:59) this step

will run before advancing to the next step by
pressing numbers or the +/- buttons on the
keypad. (To DISABLE the timer for this
step, set the timer to 0:00
.)


NOTE: If ‘DFIB’ is set to ‘N’ and ‘TMR’ is set to 0:00,
the scenario will not automatically advance to the
next step from this step
. This will allow you to decide
when to advance to the next step manually by pressing
the ‘+’ button on the keypad.

To end a running scenario, either repeatedly press
the ‘+’ button to advance through each step or press the
‘Run’ button to end it immediately.

9. Move the cursor to the ‘WAVEFORM’ position

10. Select the waveform number (0 through 80) to

use for this scenario step.


To END SCENARIO, select waveform 0.

NOTE: Be sure to end each scenario with waveform 000.
If ‘END SCENARIO’ is selected as step number 1, the
scenario will not appear to run, since this waveform will
tell it to end as soon as it starts.

If you have five (5) steps in your scenario, and want to
change it to a three (3) step scenario, and you do not
select
‘END SCENARIO’ as the waveform for step 4,
instead of stopping after step 3, the scenario will
continue with step 4 since it was not changed to ‘END
SCENARIO.’ It will then finish step 4 and continue on
with step 5, since the data was not removed from the
scenario.

11. Either press and hold the ‘PROG>’ button for five

(5) seconds to save your changes or start the
process again by selecting a different scenario or
step number. Before saving, the unit will prompt
the user to verify the edits should be saved (+ for
yes and for no).