2 technical data, 2technical data – Nor-Cal CDG 160 OP Lit User Manual

Page 10

Advertising
background image

10

CDG160-OP-LIT

(2010-06)

2

Technical Data

Measurement range

→ "Validity"

Accuracy

1

0.4% of reading

)

Temperature effect on zero

0.005% F.S./ °C

Temperature effect on span

0.02% of reading / °C

Resolution

0.003% F.S.

Gas type dependence

none

Output signal analog
(measurement signal)

Measurement range

0 … +10 V

Voltage range

–5 … +10.24 V
(limited to +10.24 V)

Relationship voltage-pressure

linear

Output impedance

0

Ω (short-circuit proof)

Loaded impedance

>10 k

Response time

2

30 ms

)

Identification

Resistance R

Ident

Voltage


13.2 k

Ω referenced to

supply common
≤5 V

1)

Non-linearity, hysteresis, repeatability in the calibrated range at 25 °C

ambient operating temperature without temperature effects after
operation of 2 h.

2)

Increase 10 … 90 % F.S.