Evacuate the gauge to a pressure according to the table

below:

F.S.

Recommended final pressure
for zero adjustment

1000 Torr
100 Torr

20 Torr

10 Torr

2 Torr

1 Torr

<5

×10

-2

Torr

<5

×10

-3

Torr

<1

×10

-3

Torr

<5

×10

-4

Torr

<1

×10

-4

Torr

<5

×10

-5

Torr

If the final pressure is too high for zero adjustment (>25%
of the F.S.), the zero cannot be reached and the
<STATUS> LED blinks green. If this is the case, activate
the factory setting and adjust the zero again (

 30).

Operate the gauge for at least 2 hours under constant

ambient conditions (until the signal is stable).

Briefly press the <ZERO> button with a pin (max.

ø1.1 mm). The zero adjustment runs automatically. The
<STATUS> LED blinks until the adjustment (duration

≤8 s)

is completed.

Press the button briefly

max. ø1.1 mm