Calibration Matrices

CalMAN is able to access any meter calibration matrices in the CRI meter’s internal user memory. As
supplied from the factory, there are calibration matrices in the factory memory, but none in the user
memory (Figure 1). Before using the meter with CalMAN, you need to copy any of the calibration
matrices that you wish to use from the factory memory to the user memory. You need to check this by
running the “Colorimeter Utility,” supplied by CRI.

Figure 1: The Calibration tab on the Colorimeter Utility software shows any calibration matrices that are loaded in the
meter's user memory.

To copy calibration matrices from the meter’s factory memory to the user memory:

1. In the Colorimeter Utility software, on the Calibration tab, click the Select Calibration Matrix

button, to the lower right (Figure 1).