Fluke 787 User Manual

Page 20

Advertising
background image

787
Users Manual

12

Table 4. Rotary Switch Positions for mA Output

No.

Position

Default Function

Pushbutton Actions

 OUTPUT

[ mA

Test leads in SOURCE:

Source 0 % mA

Test leads in

SIMULATE:

Sink 0 % mA

% STEP X or W: Adjusts output up or down to the next 25 % step

COARSE Z or Y: Adjusts output up or down 0.1 mA

FINE X or W: Adjusts output up or down 0.001 mA

OUTPUT
mA
J

Test leads in SOURCE:

Source repeating
0 % -100 %-0 %
slow ramp (E)
Test leads in

SIMULATE:

Sink repeating
0 % -100 %-0 %
slow ramp (E)

BLUE cycles through:

Fast repeating 0 % -100 % - 0 % ramp (P on display)

Repeating 0 % -100 % - 0 % ramp in 25 % steps (N on display)

Slow repeating 0 % -100 % - 0 % ramp (E on display)

Advertising