Fluke 787 User Manual

Page 47

Advertising
background image

ProcessMeter™

Specifications

39

AC Current Measurement

Range

45 Hz to 2 kHz

Resolution Accuracy,

±

(% of Reading + Counts)

Typical Burden

Voltage

1.000 A (Note)

0.001 A

1 % + 2

1.5 V/A

Note: 440 mA continuous, 1 A 30 seconds maximum on, 5 minutes off

Specifications are valid from 5 % to 100 % of amplitude range.
AC conversion: true rms
Maximum crest factor: 3
For non-sinusoidal waveforms, add

±

(2 % reading + 2 % f.s.) typical

Overload protection 440 mA, 1000V fast-blow fuse

DC Current Measurement

Range Resolution

Accuracy,

±

(% of Reading + Counts)

Typical Burden

Voltage

30.000 mA

0.001 mA

0.05 % + 2

14 mV/mA

1.000 A (Note)

0.001 A

0.2 % + 2

1.5 V/A

Note: 440 mA continuous, 1A 30 seconds maximum on, 5 minutes off

Overload protection: 440 mA, 1000V fast-blow fuse