Auto ramping the ma output – Fluke 787 User Manual

Page 35

Advertising
background image

ProcessMeter™

Using the Current Output Functions

27

Table 10. mA Step Values

Step

Value (for each span setting)

4 to 20 mA

0 to 20 mA

0 %

4.000 mA

0.000 mA

25 %

8.000 mA

5.000 mA

50 %

12.000 mA

10.000 mA

75 %

16.000 mA

15.000 mA

100 %

20.000 mA

20.000 mA

125 %

24.000 mA

120 %

24.000

mA

Auto Ramping the mA Output

Auto ramping gives you the ability to continuously apply a
varying current stimulus from the Meter to a transmitter,
while your hands remain free to test the response of the
transmitter. Select either sourcing or simulating by
choosing the SOURCE or SIMULATE jacks.

When the rotary switch is in the OUTPUT mA J
position, the Meter produces a continuously repeating 0 %
- 100 % - 0 % ramp in your choice of three ramp
waveforms:

E

0 % - 100 % - 0 % 40-second smooth ramp,

(default)

P

0 % - 100 % - 0 % 15-second smooth ramp

N

0 % - 100 % - 0 % Stair-step ramp in 25 % steps,

pausing 5 seconds at each step. Steps are listed

in Table 10.

The ramp times are not adjustable. Press the BLUE
pushbutton to cycle through the three waveforms.

Note

At any time during auto ramping, you can freeze
the ramp simply by moving the rotary switch to
the
[ mA position. Then you can use the
COARSE, FINE, and % STEP adjust pushbuttons
to make adjustments.