28

Power-Up Options

To select a power-up option, hold down the pushbutton
shown in Table 11 while turning the rotary switch from
OFF to any on position. Wait 2 seconds before you
release the pushbutton after powering up the Meter. The
Meter beeps to acknowledge the power-up option.

Only the setting for current span is retained when the
power is turned off. The others have to be repeated for
each operating session.

You may activate more than one power-up option by
holding down more than one pushbutton.

Table 11. Power-Up Options

Option Pushbutton

Default Action

Taken

Change current span 0 % setting

K Remembers

last setting

Toggles between 0 and 4 mA

Disable beeper

T Enabled

Disables

beeper

Disable auto power-off

BLUE

Enabled

Disables the feature that turns off the Meter
power after 30 minutes of inactivity. Auto
power off is disabled regardless of this
option if MIN MAX recording is on.