Fluke 787 User Manual

Page 46

Advertising
background image

787
Users Manual

38

DC Millivolts Measurement

Range (mV dc)

Resolution

Accuracy (% of Reading + Counts)

400.0

0.1 mV

0.1 % + 1

AC Volts Measurement

Range (ac)

Resolution

Accuracy,

±

(% of Reading + Counts)

50 Hz to 60 Hz

45 Hz to 200 Hz

200 Hz to 500 Hz

400.0 mV

0.1 mV

0.7 % + 4

1.2 % + 4

7.0 % + 4

4.000 V

0.001 V

0.7 % + 2

1.2 % + 4

7.0 % + 4

40.00 V

0.01 V

0.7 % + 2

1.2 % + 4

7.0 % + 4

400.0 V

0.1 V

0.7 % + 2

1.2 % + 4

7.0 % + 4

1000 V

1 V

0.7 % + 2

1.2 % + 4

7.0 % + 4

Specifications are valid from 5 % to 100 % of amplitude range.
AC conversion: true rms
Maximum crest factor: 3
For non-sinusoidal waveforms, add

±

(2 % reading + 2 % f.s.) typical

Input impedance: 10 M

Ω

(nominal), < 100 pF, ac-coupled

Common mode rejection ratio: >60 dB at dc, 50 Hz, or 60 Hz