Using the input terminals – Fluke 289 User Manual

Page 21

Advertising
background image

True-rms Digital Multimeters

Features

11

Using the Input Terminals

All functions except current use the W and COM input
terminals. The two current input terminals (A and mA/

μ

A) are

used as follows:

Current from 0 to 400 mA, use the

and COM terminals.

Current between 0 and 10 A use the A and COM terminals.

est04.emf

Figure 4. Input Terminals

Table 5. Input Terminals

Terminal Description

A

Input for 0 A to 10.00 A current (20VA
overload for 30 seconds on, 10 minutes off),
frequency, and duty-cycle measurements.

Input for 0 A to 400 mA current
measurements, frequency, and duty cycle.

COM

Return terminal for all measurements.

Input for voltage, continuity, resistance,
diode test, conductance, capacitance,
frequency, temperature, period, and duty-
cycle measurements.