Min max, recording, and peak specifications, 76 min max, recording, and peak specifications – Fluke 289 User Manual

Page 86

Advertising
background image

287/289
Users Manual

76

MIN MAX, Recording, and Peak Specifications

Function Nominal

Response

Accuracy

200 ms to 80% (dc function)

Specified accuracy ±12 counts for changes >425 ms in duration in manual
range.

MIN MAX, Recording

350 ms to 80 % (ac function)

Specified Accuracy ±40 counts for changes >1.5 s in duration in manual
range.

Peak

250

μ

S (peak)

[1]

Specified accuracy ±100 counts

[2]

up to 5,000 count (full range) reading.

For higher peak reading (to 12,000 counts), specified accuracy ±2 %

[3]

of

reading.

Crest Factor

350 ms to 80 %

For periodic waveforms from 50 to 440 Hz ± (4 % + 1 count).

[1]

For repetitive peaks; 2.5 ms for single events. Peak not specified for 500 µA DC, 50 mA DC, 5 A DC.

[2]

200 counts in 500 mV AC, 500

μ

A AC, 50 mA AC, 5 A AC.

[3]

3 % in 500 mV AC, 500

μ

A AC, 50 mA AC, 5 A AC.