Dc current specifications, 71 dc current specifications – Fluke 289 User Manual

Page 81

Advertising
background image

True-rms Digital Multimeters

Detailed Specifications

71

DC Current Specifications

Accuracy

AC over DC, DC over AC, AC + DC

[1]

Function Range Resolution

DC

[1][3]

20 to 45 Hz

45 Hz to 1 kHz

1 to 20 kHz

20 to 100 kHz

[5]

500

μ

A 0.01

μ

A

0.075 % + 20

1 % + 20

0.6 % + 20

0.6 % + 20

5 % + 40

DC

μ

A

[4]

5000

μ

A 0.1

μ

A

0.075 % + 2

1 % + 5

0.6 % + 5

0.6 % + 10

5 % + 40

50 mA

0.001 mA

0.05 % + 10

[6]

1 % + 20

0.6 % + 20

0.6 % + 20

5 % + 40

DC mA

[4]

400 mA

0.01 mA

0.15 % + 2

1 % + 5

0.6 % + 5

1.5 % + 10

5 % + 40

5 A

0.0001 A

0.3 % + 10

1.5 % + 20

0.8 % + 20

3 % + 40

[5]

Not

Spec’d

DC A

[2]

10 A

0.001 A

0.3 % + 2

1.5 % + 10

0.8 % + 10

3 % + 10

[5]

Not

Spec’d

[1]

AC+DC ranges are specified from 2 % to 140 % of range.

[2]

20 A for 30 seconds on, 10 minutes off. >10 A not specified.

[3]

Add 20 counts in dual display ac over dc, dc over ac or ac+dc.

[4]

400 mA continuous; 550 mA for 2 minutes on, 1 minute off.

[5]

Verified by design and type tests.

[6] Temperature

coefficient:

0.1 X (specified accuracy)/

°

C (<18

°

C or > 28

°

C)