Abbott Freestyle Freedom Lite User Manual

Page 12

Advertising
background image

17

18

Message

What It Means

What To Do

There may be a problem

with the test strip

Meter error

Very high blood glucose

higher than 27.8 mmol/L

HIGH control solution

was applied when the

temperature was too

cold. (Applies only to

control solution labeled

as HIGH).

1. Follow your healthcare professional’s advice to

treat high blood glucose if you have symptoms

such as:

Fatigue

Thirst

Excess urination

Blurry vision

2. Perform a control solution test using a new

test strip. If the test results are within the range

printed on the test strip vial, retest using blood

and a new test strip.

3. If the control solution result is out of range or the

error reappears, contact Customer Care.

Message

What It Means

What To Do

Incorrect test procedure.

For example, putting

blood on the test strip

before inserting it into

the meter or applying
blood before

and

appear

There may be a problem

with the test strip

Meter error

1. Wait for

and to appear before applying

blood or control solution.

2. Perform a control solution test using a new

test strip. If the test results are within the range

printed on the test strip vial, retest using blood

and a new test strip.

3. If the control solution result is out of range or the

error reappears, contact Customer Care.

There may be a problem

with the test strip

Meter error

1. Perform a control solution test using a new

test strip. If the test results are within the range

printed on the test strip vial, retest using blood

and a new test strip.

2. If the control solution result is out of range or the

error reappears, contact Customer Care.

www.AbbottDiabetesCare.com

ART22311-101_rev-B.indd 17-18

8/28/14 11:30 AM

Advertising