Performing a control solution test, Performing a control solution test 1 – Abbott Freestyle Freedom Lite User Manual

Page 15

Advertising
background image

23

24

Performing a Control Solution Test

1.

Check control solution expiry date.

IMPORTANT: Do not use control solution past expiry date. Discard

control solution 3 months after opening or on the expiry date printed on

the bottle whichever comes first. (Example: open April 15, discard July

15; write the discard date on the side of the bottle.)

2.

Remove test strip from vial.

3.

Insert test strip to turn meter on.

Note: The meter turns off after 2 minutes of inactivity.

Remove and reinsert the unused test strip to restart

the meter.

If the meter does not turn on go to Troubleshooting section.

Top

Printed side up

Sample areas

4.

Confirm System Check Screen.

This display appears when you turn on the meter. Do not use the meter if

the display check screen does not exactly match the example. Contact

Customer Care.

See The FreeStyle Freedom Lite Meter and How It Works section for more

information.

If the system check screen does not appear, go to Troubleshooting section.

5.

Apply and hold control solution to 1 sample area of the test strip until:

You see short lines moving clockwise on the display or

You hear a beep

This indicates the test strip obtained enough control solution and the meter is

processing. If the test does not start after applying the control solution, go to

Troubleshooting section.

IMPORTANT: If a short line does not appear after 5 seconds, the

sample may be too small. Do not add control solution to both sides

of the strip. The test strips are for single use only. Do not reuse test

strips. Dispose of used test strips.

www.AbbottDiabetesCare.com

ART22311-101_rev-B.indd 23-24

8/28/14 11:30 AM

Advertising