Important Test Strip Information

Please read the FreeStyle Lite test strip package insert for more information.

•

Use the test strips within the meter’s operating range: 40° – 104° F (4° – 40° C).

•

Store test strips in a cool, dry place between 40° – 86° F (4° – 30° C).

•

Store test strips away from direct sunlight and heat.

•

Store test strips in their original vial only.

•

Clean and dry your hands before removing a strip from the vial.

•

Close the vial cap tightly immediately after removing a test strip.

•

Do not transfer test strips to another container.

•

Do not store test strips outside the vial.

•

Do not bend, cut or alter test strips.

•

Do not apply blood to both sides of the test strip.

•

Do not press the test strip against the test site.

•

Do not scrape the blood onto the test strip.

•

Do not apply blood to the flat side of the test strip.

•

Do not apply blood to the test strip when the test strip is out of the meter.

•

Do not put blood or foreign objects into the meter.

Warning: Drying agents in the test strip vial or cap may be harmful if inhaled or

swallowed and may irritate skin and eyes.

Performing a Blood Glucose Test

1.

Remove test strip from vial.

2.

Insert test strip to turn meter on.

Note: The meter turns off after 2 minutes of inactivity.

Remove and reinsert the unused test strip to restart

the meter.

If the meter does not turn on, go to Troubleshooting section.

3.

Confirm System Check Screen.

•

This display appears when you turn on the meter. Do not use the meter if

the display check screen does not exactly match the example. Contact

Customer Care.

•

See The FreeStyle Freedom Lite Meter and How It Works section for more

information.

•

If the system check screen does not appear, go to Troubleshooting section.

4.

Obtain a blood drop.

When the

and appears, use the lancing device to obtain a blood drop.

(See FreeStyle lancing device insert for instructions.)

The test strip requires only a 0.3 microliter of blood for accurate results.