Troubleshooting – Abbott Freestyle Freedom Lite User Manual

Page 13

Advertising
background image

19

20

Troubleshooting

What It Means

What To Do

1. The meter

does not

enter test

mode after

inserting a

test strip.

Test strip is inserted

upside down, or bottom

end in, or not fully

inserted into the meter

1. Insert test strip print side up, top end in.

No battery is installed

Battery is installed

incorrectly

1. Install battery with (+) facing up.

Dead battery

1. Replace battery. Reset date and time, if

necessary.

2. If meter still does not enter test mode,

contact Customer Care.

Defective test strip or

meter

Blood or foreign objects

put in the test strip port

1. Contact Customer Care.

What It Means

What To Do

2. The test

does not

start after

applying

the blood

drop.

Blood drop is too small

1. Add blood to the same sample area within

60 seconds, or repeat the test using a new

test strip and a larger blood drop.

Defective test strip

Sample applied after

meter turns off

1. Repeat the test using a new test strip. Wait for

and to appear before applying blood

or control solution.

2. If problem persists, contact Customer Care.

Defective meter

1. Contact Customer Care.

www.AbbottDiabetesCare.com