21

22

FreeStyle Control Solution

Purpose

FreeStyle control solution is a red liquid that contains a fixed amount of glucose.
The purpose of doing a control solution test is:

To practise testing without using your own blood

To confirm that the meter and test strips work together properly

When you are unsure of your blood glucose test results

Contact Customer Care for information on how to obtain control solution.

Control Solution Information

Use only FreeStyle control solution with the meter.

Replace the cap securely on the bottle immediately after use.

Control solution results are accurate only between 59° – 104° F (15° – 40° C).

Do not add water or other liquid to control solution.

There are three levels of control solution available (Low, Normal, High).

For information on how to obtain control solution, contact Customer Care.

IMPORTANT: Control solution results should fall within the control

solution range printed on the test strip vial. Do not use this range

when testing blood glucose.

Out of Range Control Solution Results

Repeat the test if control solution results are outside the range printed on

the test strip vial.

Stop using the meter if control solution results are consistently outside the

range printed on the test strip vial. Contact Customer Care.

Causes of out-of-range results include:

Expired or bad control solution

Expired or bad test strip

Testing error

Watered-down control solution

Meter malfunction

Test strip deterioration

Control solution test done outside 59° – 104° F (15° – 40° C)

www.AbbottDiabetesCare.com

ART22311-101_rev-B.indd 21-22

8/28/14 11:30 AM

Advertising