13

Programming (Customizing) the Model 300 for Testing


From the system prompt,

-.-, press the Shift/Cancel button until the yellow SETUP indicator

is on, and “SET UP” is displayed.

The yellow button legend becomes the active legend to follow for the buttons.

Use the

and buttons to scroll through the set up parameters. Once you have reached a

parameter that you want to change, press

SELECT to edit its value. EDIT OPT is

momentarily displayed. Then use the

and buttons to change the value. When done

changing the value, press

SAVE to save the new value (SAVE OPT is momentarily

displayed) or

CANCEL to exit without saving.

The set up parameters are as follows:

TDLY 0

Test Delay: Set a delay of 0 to 15 seconds after the button is pressed
(for a test) before the test begins. The factory default is 0 seconds.

GPLT DLY

Choose an option for grounding the plate:

GPLT DLY

ground the plate during the test delay

GPLT <DLY ground the plate prior to the test delay, but not

during the test delay

GPLT DLY> ground the plate after the test delay, but not during

the test delay

The factory default is to ground the plate during the test delay.

VSTP 100

Decay Stop Voltage: Set the desired stop voltage. The stop voltage
can be set to any value from 10V to 990V in 1V increments. The
factory default is 100V.

DLEN 30

Decay Test Length: Set the desired test length for a decay test before
test “times out” and ends the decay test. The decay test length can be
set to any value from 10 to 999 seconds in 1-second increments. The
factory default is 30 seconds.

BLEN 30

Balance (Offset Voltage) Test Length: Set the desired test length for a
balance test before the test “times out” and terminates the balance
test. The balance test length can be set to any value from 0 to 9999
seconds in 1-second increments. Choosing zero (0) as a test length
causes the balance test to run forever. The factory default is 30
seconds.

PKDL 3

Balance Peak Delay: Set the desired time delay after the start of a
balance test before the peak detectors begin to record peak offset
voltages. The balance peak delay can be set to any value from 10 to
300 seconds in 1-second increments. The Balance Peak Delay must
be less (shorter) than the balance test length. The factory default is 3
seconds.

PWR 300

DC Auto Power Off: Set the desired time that the Model 300 will
operate on battery before shutting off. The DC Auto Power Off can be
set to any value from 10 to 9999 seconds in 1-second increments.
Choosing zero (0) as a time causes the Model 300 not to automatically
shut off. The factory default is 300 seconds.