APPENDIX V

Error Messages



Error Messages on RECALL test data:
----------------------------------

+DCY CFAI

+decay charge failed.

-DCY CFAI

-decay charge failed.
The Model 300 charged for 30 seconds and the plate
voltage did not reach to 1100v.


+DCY DFAI

+decay failed.

-DCY DFAI

-decay failed.
It is caused by one of the following:

1)

After charging to 1100v, the decay test ran for 1

minute without getting to 1000v, and the decay
timer did not start.

2) After the decay timer started at 1000v, the decay

test ran to the maximum decay length and the
voltage did not reached to the V stop setting
(timed out).


+AVG FAIL
-AVG FAIL

The decay average fails to calculate on an auto
sequence test when one or more decay tests fails.



Error Message during operation:
-------------------------------

COMM ERR

The plate isn't connected to the base unit. Check
connection.


CHRG CFAI

Charge failed.
On a decay test, the Model 300 charged for 30
seconds without getting to 1100v.


DCAY DFAI

Decay failed. Refer to above "DCY DFAI" for the
cause.


BAD BAT

Bad battery

When the Model 300 is plugged in to AC power, the
battery charger starts. If charging for 5 minutes
and the battery voltage is still below 4v, then the
Bad Battery flag is set and the charger stops
charging until the flag is cleared. The device is