AMETEK MX CTSL User Manual

Page 24

Advertising
background image

User Manual

MX-CTSL Compliance Test System

California Instruments

Revision J

24


2.8

EN 61000-4-14 Voltage Fluctuations

The EN 61000-4-14 is aimed at testing a product for immunity against voltage fluctuations that can occur
on the public utility. The nature of the voltage fluctuations is different from those tested under EN 61000-4-
11 however. The fluctuations covered in this standard are less severe and involve typical high and low line
conditions.

California Instruments’ CTS system supports full compliance EN 61000-4-14 testing. The EN 61000-4-14
test is implemented in the MXGUI Windows

→

software. To accommodate changes in the specification,

test levels and duration can be changed by the operator or loaded from disk if needed.

2.9

EN 61000-4-17 DC Ripple

The EN 61000-4-17 is aimed at testing a DC powered product for immunity against voltage ripple. Ripple is
often caused by feed-through from the AC input section of a DC supply and exhibits itself as a small AC
signal riding on the DC output. The EN 61000-4-17 standards covers test levels and frequencies for testing
the immunity of DC products against such phenomena.

The CTS system supports full compliance EN 61000-4-17 testing The EN 61000-4-17 test is implemented
in the MXGUI Windows

→

software. To accommodate changes in the specification, test levels and

duration can be changed by the operator or loaded from disk if needed.

Advertising