SLH-Series AC High Power Electronic Load

GPIB/RS232 Programming Operation

M540073-01 Rev C

4-9

WATT Meter ON/OFF

Purpose:
To set display of power meter of AC series electronic load.
Command Syntax:
WATT{SP}{0|1|OFF|ON}{;|NL}
Description:
This command is to set the power meter display. This command must be used in
conjunction with PRES:OFF. When setting to ON, the top monitor will change from voltmeter
to Watt meter while the monitor at the bottom will change from ammeter to Volt-Ameter (VA)
and the unit is “W” and “VA” respectively. When setting to OFF, the Watt meter on the top
will change back to voltmeter while the VA meter at the bottom will change back to ammeter
and the unit is “Vrms” and “Arms” respectively.
Example:
PRES OFF
WATT ON ; to display WATT, VA meter.
WATT OFF ; to display Voltage, Current meter.

WAVEFORM BANK

Purpose:
To set waveform bank for AC electronic load modules.
Command Syntax:
BANK{SP}{d}{;|NL} d:0-10
Description:
This command is to set the desired waveform bank.
1. waveform bank 0-4 are sine wave.
2. waveform bank 5-9 are square wave.
3. waveform bank 10 is DC.
4. There are five (5) waveforms for each of eleven (11) waveform banks for a total of 55
waveforms. Waveform information is shown in Table 4-10.

Example:
BANK 1, to set waveform bank 1.
BANK 10, to set waveform bank 10.