Operation

Sorensen SLM-Series DC Load

3-2

M540071-01 Rev C

1 Module

Label

Shows the load model name and its maximum ratings.

2 NG

LED

Lights to indicate "No Go" (fail) when Vmeter, Ameter, or Wattmeter exceeds the upper or
lower limit set. Off, not lit, indicates "Go" (no fail).

3

CC, CR, CV, CP LEDs

and MODE Key

•

LEDs indicate which mode is in operation:

◦ CC = Constant Current
◦ CR = Constant Resistance
◦ CV = Constant Voltage
◦ CP = Constant Power

•

The MODE key switches from one mode to the next.

4 REM

LED

Indicates remote operation:

•

When lit, the unit is under remote control and cannot be operated through the front

panel keys. (Remote controller releases control by GPIB command).

•

When off, the unit is under local control and can be manually operated using the front
panel keys.

5

Upper 5-Character Digital Meter (DM) and Associated LEDs

The DM is for multi-purpose display, depending on selected mode:

•

In PRESet OFF mode, displays voltage of DC input terminal or V

SENSE

BNC input.

•

In PRESet ON mode, displays values being programmed for CV, CR, or CP (value for

CC is displayed in lower DM).

•

In WATT On mode, displays load Watt meter.

•

In LIMIT ON mode:

◦ displays GO/NG upper limit values for voltage, current, and power
◦ displays SENSE to indicate V

SENSE

control

◦ displays voltage settings for LOAD ON or for LOAD OFF

Associated LEDs indicate the displayed value's unit:

•

V = Value displayed is in Volts

•

A = Value displayed is in Amps

•

Ω = Value displayed is in Ohms

•

W = Value displayed is in Watts

•

ms = Value displayed is in milliseconds

•

A/

μs (or mA/μs) = Value displayed is in Amps per microsecond (or milliamps per

microsecond)