Specifications – Beurer BC 16 User Manual

Page 21

Advertising
background image

21

9. specifications

Model no.

BC 16

Measurement

method

Oscillometric, non­invasive blood

pressure measurement on the wrist

Measurement range

Cuff pressure 0 – 299 mmHg,

systolic 30 – 260 mmHg,

diastolic 30 – 260 mmHg,

Pulse 40 – 199 beats/minute

Display accuracy

Systolic ± 3 mmHg,

diastolic ± 3 mmHg,

pulse ± 5 % of the value shown

Measurement

inaccuracy

Max. permissible standard deviation

according to clinical testing:

systolic 8 mmHg /diastolic 8 mmHg

Memory

2 x 50 memory spaces

Dimensions

L 82 mm x W 69 mm x H 28 mm

Weight

Approx. 120 g (without batteries)

Cuff size

135 to 220 mm

Permissible operating

conditions

+10 °C to +40 °C, 40 – 85 % relative

air humidity (non­condensing)

Permissible storage

conditions

­10 °C to +60 °C, 10 – 90 % relative

air humidity, 800­1050 hPa ambient

pressure

Power supply

2 x 1,5 V

AA batteries

Battery life

For approx. 300 measurements,

depending on the blood pressure

level and/or pump pressure

Accessories

Instruction for use, 2 x 1.5 V AA bat­

teries, storage box

Classification

Internal supply, IPX0, no AP or

APG, continuous operation, type BF

applied part

Technical information is subject to change without notifica­

tion to allow for updates.
• This unit is in line with European Standard EN 60601­1­2

and is subject to particular precautions with regard to

electromagnetic compatibility (EMC). Please note that

portable and mobile HF communication systems may

interfere with this unit. More details can be requested from

the stated Customer Service address or found at the end

of the instructions for use.

• This device is in line with the EU Medical Devices Direc­

tive 93/42/EC, the “Medizinproduktegesetz” (German

Medical Devices Act) and the standards EN 1060­1

(non­invasive sphygmomanometers, Part 1: General

requirements), EN 1060­3 (non­invasive sphygmoma­

nometers, Part 3: Supplementary requirements for elec­