Electromagnetic Compatibility Information

Table 1

For all ME EQuIPMEnT and ME sYsTEMs

Guidance and manufacture’s declaration – electromagnetic emissions
The BC 16 is intended for use in the electromagnetic environment specified below.

The customer or the user of the BC 16 should assure that it is used in such an environment.
Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1

The BC 16 uses RF energy only for its internal function. Therefore, its RF emissions are

very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class B

The BC 16 is suitable for use in all establishments, including domestic establish­

ments and those directly connected to the public low­voltage power supply network

that supplies buildings used for domestic purposes.

Harmonic emissions IEC 61000­3­2

Not applicable

Voltage fluctuations/flicker emissions

IEC 61000­3­3

Not applicable

Table 2

Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 16 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 16 should assure that it

is used in such an environment.
Immunity test

IEC 60601 test level Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000­4­2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered

with synthetic material, the relative humidity should be at least 30 %.

Electrical fast transient/burst

IEC 61000­4­4

± 2 kV for power

supply lines

± 1 kV for input/

output lines

± 2 kV for power

supply lines

N/A

Mains power quality should be that of a typical commercial or hos­

pital environment.

Power frequency (50/60 Hz)

magnetic field IEC 61000­4­8 3 A/m

3 A/m

Power frequency magnetic fields should be at levels characteristic

of a typical location in a typical commercial or hospital environment.

noTE: U

T

is the a.c. mains voltage prior to application of the test level.