Electromagnetic Compatibility Information

Table 1

Guidance and manufacturer’s declaration – electromagnetic emissions
The BC 32 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 32 should assure that it is

used in such an environment.
Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1

The BC 32 uses RF energy only for its internal function. Therefore, its RF emissions

are very low and are not likely to cause any interference in nearby electronic equip-

ment.

RF emissions CISPR 11

Class B

The BC 32 is suitable for use in all establishments, including domestic establish-

ments and those directly connected to the public low-voltage power supply network

that supplies buildings used for domestic purposes.

Harmonic emissions IEC 61000-3-2

Not applicable

Voltage fluctuations/flicker emissions

IEC 61000-3-3

Not applicable

Table 2

Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 32 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 32 should assure that it is

used in such an environment.
Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are

covered with synthetic material, the relative humidity should be

at least 30 %.

Power frequency (50/60 Hz)

magnetic field IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at levels characteris-

tic of a typical location in a typical commercial or hospital envi-

ronment.