Beurer BC 32 User Manual

Page 75

Advertising
background image

75

Table 3

Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 32 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 32 should assure that it is

used in such an environment.
Immunity test IEC 60601 test

level

Compliance

level

Electromagnetic environment – guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 V

rms

150 kHz

to 80 MHz

3 V/m 80 MHz

to 2,5 GHz

Not applicable

3 V/m

Portable and mobile RF communications equipment should be used no closer to any part

of the BC 32, including cables, than the recommended separation distance calculated from

the equation applicable to the frequency of the transmitter.
Recommended separation distance:

d = 1,2

80 MHz to 800 MHz

d = 2,3

800 MHz to 2,5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to

the transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site

survey,

a

should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,

objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,

AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment

due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the

BC 32 is used exceeds the applicable RF compliance level above, the deivce should be observed to verify normal operation. If abnormal per-

formance is observed, additional measures may be necessary, such as reorienting or relocating the BC 32.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Advertising