Electromagnetic Compatibility Information

Table 1

Guidance and manufacturer’s declaration – electromagnetic emissions
The BM 19 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 19 should assure

that it is used in such an environment.
Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1

The BM 19 uses RF energy only for its internal function. Therefore, its RF emissi­

ons are very low and are not likely to cause any interference in nearby electronic

equipment.

RF emissions CISPR 11

Class B

The BM 19 is suitable for use in all establishments, including domestic establish­

ments and those directly connected to the public low­voltage power supply net­

work that supplies buildings used for domestic purposes.

Harmonic emissions IEC 61000­3­2

Not applicable

Voltage fluctuations/flicker emissions

IEC 61000­3­3

Not applicable

Table 2

Guidance and manufacturer’s declaration – electromagnetic immunity
The BM 19 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 19 should assure

that it is used in such an environment.
Immunity test

IEC 60601 test level Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000­4­2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are

covered with synthetic material, the relative humidity should

be at least 30 %.

Power frequency (50/60 Hz)

magnetic field IEC 61000­4­8 3 A/m

3 A/m

Power frequency magnetic fields should be at levels characte­

ristic of a typical location in a typical commercial or hospital

environment.