Cleaning and storing the instrument, Error messages / troubleshooting, Specifications – Beurer BM 35 User Manual

Page 19

Advertising
background image

19

display

). If you continue pressing the MEM button, the latest

individual results are displayed with date and time.

• To clear the memory, press the MEM button, then the dis-

play shows

. Press the + button to select the user mem-

ory and confirm by pressing MEM again. Press and hold the
+ and

keys simultaneously for 5 seconds (the display will

show

).

• If you wish to change the user memory, see “Select memory”.

7. Cleaning and storing the instrument

• Clean your device and cuff carefully only with a slightly mois-

tened cloth.

• Do not use detergents or solvents.
• On no account must you immerse the computer in water,

otherwise liquid can enter it and cause demage.

• When storing the device, make sure that no heavy objects

are placed on top of it. Remove the batteries. The cuff tube
should not have any sharp kinks

8. Error messages / troubleshooting

In case of faults, the _ message appears in the display.
Error messages can occur when
1. inflation pressure is higher than 300 mmHg,
2. blood pressure is unusually high or low

,

3. you move or talk during the measurement (in addition to

,

the cardiac arrhythmia icon

appears in the display

),

4. the cuff tube is not inserted properly

,

5. the cuff is inflated for longer than 25 seconds

.

In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do

not move or talk. Re-insert the batteries if necessary, or else
replace them.

9. Specifications

Model no.

BM 35

Measurement
method

Oscillometric, non-invasive blood pres-
sure measurement on the upper arm

Measurement range

Cuff pressure 0 – 300 mmHg,
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
Pulse 40 – 199 beats/minute

Display accuracy

Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown

Measurement inac-
curacy

Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg

Memory

2 x 60 memory spaces

Dimensions

L 135 mm x W 105 mm x H 53 mm

Weight

Approx. 170 g (without batteries)

Cuff size

23 to 33 cm

Permissible operating
conditions

+10 °C to +40 °C, ≤ 85 % relative air
humidity (non-condensing)

Permissible storage
conditions

- 20 °C to +70 °C, ≤ 85 % relative air
humidity, 700 –1060 hPa ambient pres-
sure

Power supply

4 x 1,5 V

AAA batteries