Table 2

For all ME EQuIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic immunity
The BM 58 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 58 should assure that

it is used in such an environment.
IMMuNITY test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile.

If floors are covered with synthetic material, the

relative humidity should be at least 30 %.

Electrical fast transient/burst

IEC 61000-4-4

± 2 kV for power supply lines

± 1 kV for input/output lines

± 2 kV for power supply lines

± 1 kV for input/output lines

Mains power quality should be that of a typical

commercial or hospital environment.

Surge

IEC 61000-4-5

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

Mains power quality should be that of a typical

commercial or hospital environment.

Voltage dips, short interrup-

tions and voltage variations

on power supply input lines

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 s

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 s

Mains power quality should be that of a typi-

cal commercial or hospital environment. If the

user of the BM 58 requires continued operation

during power mains interruptions, it is recom-

mended that the BM 58 be powered from an

uninterruptible power supply or a battery.

Power frequency

(50/60 Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at

levels characteristic of a typical location in a

typical commercial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.