Transferring measurements, Error messages/trouble-shooting – Beurer BM 58 User Manual

Page 27

Advertising
background image

27

8. Transferring measurements

Connect the blood pressure monitor to your PC using the

USB cable.

No data transfer may be launched whilst performing a

measurement.

PC is shown on the display. Begin the data transfer

in the “HealthManager” PC software. During the

data transfer, an animation is shown on the

display. A successful data transfer is displayed

as in figure 1. If the data transfer is unsuccessful,

an error message appears as in figure 2. In this

case, interrupt the PC connection and start the

data transfer again.
After 30 seconds of not being in use or if com-

munication with the PC is interrupted, the blood

pressure monitor switches itself off automati-

cally.

9. Error messages/trouble-shooting

In case of faults, the _ message appears in the display.

Error messages may appear if:

• systolic or diastolic pressure could not be measured (

or

appears on the display)

• systolic or diastolic pressure was outside the measure-

ment range ( or

Lo appears on the display)

• the cuff is fastened too tightly or loosely (

or

appears on the display)

• the pump pressure is higher than 300 mmHg (

appears on the display)

• pumping up takes longer than 160 seconds (

appears

on the display)

• there is a system or device error (

,

,

or

appears on the display)

• the batteries are almost empty

.

In such cases, repeat the measurement. Ensure you do not

move or speak.

If necessary, re-insert or replace the batteries.

Technical alarm – description

Should the recorded blood pressure (systolic or diastolic) lie

outside the limits specified in the section “Technical speci-

fications”, the technical alarm will appear on the display

indicating either “ ” or “

Lo”. In such cases, you should seek

medical assistance and check the accuracy of your proce-

dure.

The limit values for the technical alarm are factory set and

cannot be adjusted or deactivated. These alarm limit val-

ues are accorded second priority under the standard IEC

60601-1-8.

Abb. 1

Abb. 2

Advertising