Beurer BM 58 User Manual

Page 120

Advertising
background image

120

Table 3

For ME EQuIPMENT and ME SYSTEMS that are not lIFE-SuPPORTING

Guidance and manufacturer’s declaration – electromagnetic immunity
The BM 58 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 58 should assure that

it is used in such an environment.
IMMuNITY

test

IEC 60601

test level

Compliance

level

Electromagnetic environment – guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 V

rms

150 kHz to

80 MHz

3 V/m

80 MHz to

2,5 GHz

3 V

rms

3 V/m

Portable and mobile RF communications equipment should be used no closer to any part of

the BM 58, including cables, than the recommended separation distance calculated from the

equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2

d = 1,2

80 MHz to 800 MHz

d = 2,3

800 MHz to 2,5 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the

transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,

a

should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-

tures, objects and people.