• Protect the device from damaging factors such as moisture, dirt, dust, blood, control solution or

water, marked temperature fluctuations, direct sunlight and extreme cold.

• Do not use the device in the vicinity of strong electromagnetic fields, radio systems or mobile tel-

ephones.

9. w

hat

if

there

are

problems

?

Display messages on batteries and blood glucose measurement

no.

Cause

Solution

LP

Empty battery.

Replace battery.

Ht

Temperature of the measuring environ-

ment, blood glucose monitor or test

strip above the permitted range.

Repeat the test using a new test strip as soon as

the measuring environment, measuring device

and test strips have reached room temperature

(+20 °C to +26 °C).

Lt

Temperature of the measuring environ-

ment, blood glucose monitor or test

strip below the permitted range.

Repeat the test using a new test strip as soon as

the measuring environment, measuring device

and test strips have reached room temperature

(+20 °C to +26 °C).

Err

Used or contaminated test strip in-

serted.

Insert an unused test strip that has not ex-

pired.

• Repeat the blood glucose measurement.

Err001 System error.

Remove battery and reinsert battery. Contact

Customer Services if the problem persists.

Err002 Insufficient blood on the test strip.

Repeat the measurement using a new test strip.

12

127

130

System error.

Contact Customer Services.

Unknown error messages.

Remove battery and reinsert battery. Contact

Customer Services if the problem persists.