Beurer GL 42 (mg/dL) User Manual

Page 26

Advertising
background image

26

Beurer GL42 mg/dL

Problem: device does not switch on

Cause

Solution

Empty battery.

Replace battery.

Incorrectly inserted or missing batteries.

Check whether the battery has been inserted

correctly (see “Removing the battery insulation

strips, replacing the battery” on page 14).

Test strip inserted incorrectly or not completely. Firmly insert the test strip into the slot on the

device with the contacts first. Make sure that

the front of the test strip is facing you (see “Test

strips” on page 12).

Device faulty.

Contact Customer Services.

Problem: the test does not start after inserting the test strip into the device and applying blood.

Cause

Solution

Insufficient blood or test strip not filled correctly. Repeat test using a new test strip and a larger

drop of blood. Please note the correct filling of

the test strip.

Faulty test strip.

Repeat the test using a new test strip.

Blood was applied while the device was

switched off.

Repeat the test using a new test strip and only

apply blood after and flash.

The basic settings of the device were changed

and the changes were not completed (see “Mak-

ing and changing basic settings” on page 14).

Remove the test strip and press the “On/Off”

button until “

” is displayed.

Repeat test.

Device faulty.

Contact Customer Services.