2 FIELD BUS INSTALLATION 

2.1 G

ENERAL

 

 
FLOW‐BUS is a 3‐wire, RS485‐based field bus communication system for parameter value exchange between digital 
Bronkhorst products. In this system each instrument / device is equipped with a micro‐controller for its own dedicated 
task  but  also  for  exchanging  parameter  value  information  with  other  instruments  /  devices  connected  to  the  same 
FLOW‐BUS system. 
 
FLOW‐BUS systems may have a minimum of 2 and a maximum of 126 connections. The maximum length for datelines 
between  the  first  and  the  last  connection  may  be  up  to  approx.  600  meters.  Longer  distances  are  only  possible  in 
combination  with  special  bus‐repeater  modules.  Each  instrument/device  connection  T‐part  (stub)  must  be  kept  as 
short as possible (maximum cable length 0.5 meter).  
 
The baud rate used to transport messages is 187500 baud or 400000* baud. 
FLOW‐BUS is a multi‐master network with a token‐ring architecture. 
 

 

*MBC3 type instruments have an additional baud rate available of 400000 baud. 

 